ABSTRACT

During the above time period, 161 patients were evaluated at the CSH as a result of traumatic injury and the device was placed on approximately 40 patients. In most patients, StO2 readings of greater than 70% were noted during the initial evaluation. No further information was collected from these patients. In 8 patients, convenience samples of StO2 data were collected along with pertinent physiologic data. In these patients, StO2 levels of below 70% tracked with hypotension, tachycardia, and clinical shock resulted in increases in StO2 after resuscitation maneuvers.